Introduction
Error code for your reference
Rossmax Fingertip Pulse Oximeter SB100 is used
to measure arterial oxygen saturation (% SpO2)
of hemoglobin and pulse rate, an important
indicator of your respiratory function. It is non-
invasive device intended for spot-check of adult
and pediatric whose age is over 3 at home,
hospital and clinics.
Attention: Consult the accompanying
documents. Please read this manual
carefully before use. Please be sure to keep
this manual.
SENSOR ERROR:
Sensor cannot be de-
tected, return the device
to your local distributor or
service centre.
MEASURE ERROR:
Signals cannot be detect-
ed, turn the device off and
measure again.
Installing Batteries
1. Use thumb to slide battery cover out
2. Insert or replace 2 “AAA” sized batteries down
with the correct electrical polarity.
You need to replace the batteries when
1. Battery icon is blinking on display.
2. The function button is pressed and nothing
appears on display.
Caution: Batteries may leak or explode if used
or disposed of improperly. Remove
batteries if the device will be stored for
long time. Do not use different types
or brands of batteries at the same time.
Do not mix fully charged and partially
charged batteries at the same time.
Attaching the lanyard
1. Insert the narrow end of the
lanyard through the holder.
2. Draw the other end of the
lanyard through the loop at
the narrow end and tighten.
How to use
1. Open the clip; press the Power On button as .
2. Information of software version appears; insert
one finger, nail side up, into the finger opening
of the pulse oximeter.
Note: if no finger insert, the device will auto shut
off after 8 seconds.
3. The pulse strength indication shows “ ”, pulse
oximeter begins its measurement as .
Note: make sure the finger is lying flat, Do not shake
and keep body steady during measurement
as .
4. Your SpO2 and pulse rate values will appear on
the screen after few seconds as .
Note: 1. Don’t remove your finger until the
measurement is completed.
2. If SpO2 and pulse rate cannot be detected, “
“ will appear on the screen as .
3. While pulse strength is low, the reading
will flicker.
Note:
1. The SpO2 sensor and photoelectric receiving
tube should be arranged in a way with the
subject’s arteriole in a position there between.
2. Make sure the optical path is free from any
optical obstacles like rubberized fabric.
3. Excessive ambient light may affect the measuring
result. It includes fluorescent lamp, dual ruby
light, infrared heater, direct sunlight and etc.
4. Strenuous action of the subject or extreme
electrosurgical interference may also affect the
accuracy.
Cautionary Notes
- This device is to be operated by trained person-
nel only.
- This device has no audible and it intended only
for spot-checking, but not medical result evalu-
ation.
- This device is designed to determine the per-
centage of arterial oxygen saturation of func-
tional hemoglobin. Factors that may degrade
pulse oximeter performance or aect the accu-
racy of the measurement include the following:
• Do not apply the pulse oximeter on the same
arm as a blood pressure cu, arterial catheter or
infusion line(s)
• Excessive light, such as sunlight or direct home
lighting.
• Not steady at the site of application (e.g. trem-
bling)
• Moisture in the device
• Improperly applied device
• Finger is too large or too small to t into the
device.
• Poor pulse quality
• Venous pulsations
• Anemia or low hemoglobin concentrations.
• Cardiogreen and other intravascular dyes
• Carboxyhemoglobin
• Methemoglobin
• Dysfunctional hemoglobin
• Articial nails or ngernail polish
• On ngers with anatomical changes, oedemas,
scars or burns.
- Using the device for long periods may cause
pain for people with circulatory disorders. Repo-
sition the device at least once every 4 hours to
allow the patient’s skin to breath and to check
patient’s condition regularly.
- Do not use the device near ammable or explo-
sive gas mixtures.
- Do not use the device during an MRI or CT scan.
- The device may not work when circulation is
reduced. Warm or rub the nger, or re-position
the device.
- This device is a precision electronic instrument
and must be repaired by qualied technical pro-
fessionals. Field repair of the device is not pos-
sible. Do no attempt to open the case or repair
the electronics. Opening the case may damage
the device and void the warranty.
- Do not overextend the device’s spring.
1. SpO2 icon
2. Pulse strength
indication
3. Pulse rate icon
7
1
2
3
4
5
6
4. Beats per minute
5. Power On Button
6. Battery icon
7. Battery compartment
Name/ Functions of each part
Joignez-vous à la conversation sur ce produit
Ici, vous pouvez partager ce que vous pensez du Rossmax SB100 Oxymètre de pouls. Si vous avez une question, lisez d’abord attentivement le mode d’emploi. La demande d’un mode d’emploi peut être effectuée en utilisant notre formulaire de contact.
répondre | Cela a été utile (0) (Traduit par google)